Tabella a-3 el, A-12 – Physio-Control LIFEPAK 15 Manuale d'uso
Pagina 230

A-12
Monitor/defibrillatore LIFEPAK 15 - Istruzioni operative
Tabella A-3 Limiti degli allarmi
SEGNO VITALE
(VS)
VALORE VS
PAZIENTE
LIMITI AMPI*
LIMITI STRETTI*
GAMMA DEI LIMITI
LIMITI PREDEFINITI**
INFERIORE SUPERIORE INFERIORE SUPERIORE INFERIORE SUPERIORE INFERIORE
SUPERIORE
Frequenza
cardiaca (FC)
Frequenza
polso (FP)
<60
-20
+35
-10
+25
30–150 100–250 50
150
60–79
-25
+40
-20
+30
80–104
-30
+40
-30
+30
(BPM)
>105
-35
+45
-25
+25
SpO
2
(%)
>90
-5
+3
-5
+3
50
90–100
85
100
<90
-5
+3
-5
+3
PS sistolica
(mmHg)
<90
-20
+35
-10
+25
30
245
50
200
90–114
-20
+35
-10
+25
115–140 -25
+35
-10
+20
>140
-25
+35
-10
+20
PS diastolica
(mmHg)
<65
-15
+25
-10
+25
12
210
20
150
65-90
-15
+15
-15
+10
>90
-15
+15
-15
+10
EtCO
2
(mmHg/%)
>40/5,3
-10/-1,3 +15/+2,0 -10/-1,3 +15/+2,0 5/0,7
70/9,2
15/2,0
50/6,6
≤40/5,3
-10/-1,3 +15/+2,0 -10/-1,3 +15/+2,0
CO
2
inspirata
(mmHg/%)
—
n.a.
+5/+0,7
n.a.
+3/+0,4
n.a.
0/0–10/
1,3
—
8/1.1
Frequenza
respiratoria (FR)
<15
-8
+8
-4
+4
5-15
10–60
5
30
≥15
-15
+15
-8
+8
PA sistolica
(mmHg)
<15
-6
+12
-4
+6
10
100
10
40
15–35
-8
+16
-6
+8
>35
-12
+16
-8
+10
PA diastolica
(mmHg)
<5
-4
+12
-4
+8
0
50
0
18
5–13
-4
+16
-6
+6
>13
-6
+16
-6
+6
PVC (mmHg)
≥9
-10
+10
-5
+5
0
25
0
15
ICP, LAP
(mmHg)
<15
-6
+6
-4
+4
0
40
0
18
≥15
-6
+8
-4
+6
*Quando si impostano gli allarmi, i numeri indicati vanno aggiunti ai valori dei segni vitali del paziente o sottratti da essi.
**I limiti predefiniti vengono stabiliti quando gli allarmi sono impostati come attivi.