Pagina 35: Replace coolant filter

Filling the Coolant Reservoir
Fill the coolant reservoir according to the following instructions.
WARNING
Never attempt to connect or disconnect RF cables while RF power is on. Radiated RF is a
potential health hazard.
WARNING
Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of
vapors and eye and skin contact.
CAUTION
The Calorimeter can overheat if operated without sufficient coolant. Failure to keep coolant
reservoir adequately filled can result in serious damage to the equipment.
CAUTION
Use only premixed coolant Bird part number 6091-120. Use of other solutions will damage
the instrument and void all warranties.
CAUTION
Do not apply RF power to the load without proper coolant flow. Failure to keep coolant at
proper flow rate can result in serious damage to the equipment.
The Calorimeter requires 525 ml (10 fluid ounces) of coolant. Fill the coolant reser-
voir as follows. Refer to figure 13.
1. Remove the power cord from the Calorimeter.
2. Remove the top cover from the Calorimeter case (if supplied).
23
Maintenance
Figure 13
Filling Coolant
Reservoir