2. Disconnect RF power cable.
3. Allow unit to run until
∆
T is less than 0.040°C.
4. Press front panel ON/OFF button.
+
Note: If the alarm sounds, check
∆
T. The alarm will stop when the conditions of
step 3 above are met.
5. Turn off the rear panel main power ON/OFF switch.
Emergency Shutdown
The Calorimeter has a built in alarm that will sound during equipment emergencies.
If an emergency occurs, shut down the Calorimeter as follows:
CAUTION
Never turn off the Calorimeter’s main power switch while RF source power is applied to
Calorimeter input. Failure to turn off RF power source can result in major equipment
damage.
1. Turn off RF power source device.
2. Disconnect RF power cable.
3. If coolant flow is less than 0.284, turn off the rear panel power switch.
4. If coolant flow is above 0.284, leave the unit running until
∆
T is less
than 0.040°C.
Bird Model 6091 Calorimeter
18