Calibration – Watson-Marlow 505Di Manuale d'uso
Pagina 26

26
Pumphead ml/rev
Recalibrate
Calibration dose
Head and Tubing
###rpm cw/ccw
press start
Wait or press Stop
when ready
Press Stop when
ready
####.#ml ####.#g
Press Enter
Pumphead ####ml/rev
Maximum ###rpm
501RL 505L 505LA
303 304
502AA 505BA
304MC 308MC
504MC 508MC
0.5 0.8 1.6 3.2
4.8 6.4 8.0mm
0.13 0,19 0.25
0.38 0.50 0.63 >
0.76 0.88 1.02
1.14 1.29 1.42 >
1.47 1.52 1.65
1.85 2.05 2.38 >
2.54 2.79mm <
Pumphead Tube bore
Maximum ###rpm
Dose Cal Manual
Network Setup
Calibrate
Head and
Tubing
Enter
Enter
Start
Enter
Stop
Enter actual
volume
dosed
Enter
ml/rev and max
speed with
choosen
pumphead.
Step
Step
Step
Step
Step
Enter
Step
Calibration
E1
L1
M1
Step
Enter pump
speed and
direction
Pump starts
calibrating. Wait
and press stop
when ready.
After 15 seconds
pump prompts
for user to stop
calibrating.
Select
pumphead and
tubing size
which is fitted to
pump
Enter
Pump confirms
selections and
maximum rpm
with pumphead
G1
H1
I1
J1
K1
F1
Step
Indication of
previous
calibration dose
(if completed)
D1
Enter