Watson-Marlow 505U Manuale d'uso
Pagina 16

16
313/314 (ml/min)
Portate
#
112 13 14 16 25 17 18
mm
0.5 0.8 1.6 3.2 4.8 6.4 8.0
"
1/50 1/32 1/16 1/8 3/16 1/4 5/16
313
55
1.7 3.9 14 55 121 198 275
220
6.6 15 57 220 485 790
1100
314
55
1.7 3.3 14 47 105 165 220
220
6.6 13 55 187 420 660 880
313
Numero massimo di testine
313/314 Peroxide/ Platinum Silicone
(0
≤ bar ≤ 0.5)
(0.5 ≤ bar ≤ 2.0)
# 112 13 14 16 25 17 18 112 13 14 16 25 17 18
mm 0.5 0.8 1.6 3.2 4.8 6.4 8.0 0.5 0.8 1.6 3.2 4.8 6.4 8.0
" 1/50 1/32 1/16 1/8 3/16 1/4 5/16 1/50 1/32 1/16 1/8 3/16 1/4 5/16
55
6 6 6 6 6 4 3 6 6 6 6 5 4 3
220
6 6 6 6 6 4 3 6 6 6 6 5 4 3
313/314 Marprene, Tygon, Neoprene, Fluorel
(0
≤ bar ≤ 0.5)
(0.5 ≤ bar ≤ 2.0)
# 112 13 14 16 25 17 18 112 13 14 16 25 17 18
mm 0.5 0.8 1.6 3.2 4.8 6.4 8.0 0.5 0.8 1.6 3.2 4.8 6.4 8.0
" 1/50 1/32 1/16 1/8 3/16 1/4 5/16 1/50 1/32 1/16 1/8 3/16 1/4 5/16
55
6 6 6 6 5 3 3 6 6 6 6 4 3 3
220
6 6 6 6 5 3 3 6 6 6 5 4 3 3