Set baud rate, Set receive to transmit delay, Report number of floats – VEGA VEGAFLEX 86 Modbus and Levelmaster protocol Coax probe Manuale d'uso
Pagina 82: Report receive to transmit delay

82
11 Appendice
VEGAFLEX 86 • Protocollo Modbus e LevelMaster
42288-IT
-130416
Set Baud Rate
Request:
Parametri
Length
Code/Data
Set Baud Rate
8 (12) characters ASCII
UuuBbbbb[b][pds]
Bbbbb[b] = 1200, 9600 (default)
pds = parity, data length, stop bit
(optional)
parity: none = 81, even = 71 (defau-
lt), odd = 71
Response:
Parametri
Length
Code/Data
Set Baud Rate
11 characters ASCII
Esempio: U01B9600E71
Modificare l'apparecchio all'indirizzo 1 impostando il baud rate 9600, parità pari, 7 bit di dati, 1 bit di
stop
Set Receive to Transmit Delay
Request:
Parametri
Length
Code/Data
Set Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
Response:
Parametri
Length
Code/Data
Set Receive to Transmit Delay
6 characters ASCII
UuuROK
Report Number of Floats
Request:
Parametri
Length
Code/Data
Set Receive to Transmit Delay
4 characters ASCII
UuuF
Response:
Parametri
Length
Code/Data
Set Receive to Transmit Delay
5 characters ASCII
UuuFn
n = number of measurement values
(0, 1 or 2)
Report Receive to Transmit Delay
Request:
Parametri
Length
Code/Data
Report Receive to Transmit Delay
4 characters ASCII
UuuR