11 Appendice
VEGAPULS 65 • Protocollo Modbus e LevelMaster
41365-IT
-121116
Assign Unit Number
Request:
Parametri
Length
Code/Data
Assign Unit Number
6 characters ASCII
UuuNnn
Response:
Parametri
Length
Code/Data
Assign Unit Number
6 characters ASCII
UuuNOK
uu = new Address
Set number of Floats
Request:
Parametri
Length
Code/Data
Set number of Floats
5 characters ASCII
UuuFn
Response:
Parametri
Length
Code/Data
Set number of Floats
6 characters ASCII
UuuFOK
Se il numero viene impostato su 0, non viene più segnalato alcun livello
Set Baud Rate
Request:
Parametri
Length
Code/Data
Set Baud Rate
8 (12) characters ASCII
UuuBbbbb[b][pds]
Bbbbb[b] = 1200, 9600 (default)
pds = parity, data length, stop bit
(optional)
parity: none = 81, even = 71 (de-
fault), odd = 71
Response:
Parametri
Length
Code/Data
Set Baud Rate
11 characters ASCII
Esempio: U01B9600E71
Modificare l'apparecchio all'indirizzo 1 impostando il baud rate 9600, parità pari, 7 bit di dati, 1 bit di