4 rotors, 1 rotor fa-45-12-17, Rotors 2.4.1 – Eppendorf Centrifuge 5427 R Manuale d'uso
Pagina 9: Rotor fa-45-12-17

Product description
Centrifuge 5427 R
English (EN)
9
2.4
Rotors
2.4.1
Rotor FA-45-12-17
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which are marked with the name of the centrifuge (e.g., 5427 R).
Rotor FA-45-12-17
Aerosol-tight fixed-angle rotor
for 12 reaction tubes
Max. g-force:
20 598 × g
Max. speed:
14 000 rpm
Max. load (adapter, tube
and contents):
12 × 9.5 g
Vessels
Vessel
Adapters
Adapter base
Max. g-force:
Capacity
Order no.
(international)
Tube diameter
Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel
–
–
20 598 × g
5 mL
Ø 17 mm
14 000 rpm
-/12
9.4 cm
Vessel
5820 768.002
open
17 530 × g
1.5 mL/2.0 mL
Ø 11 mm
14 000 rpm
1/12
8.0 cm
HPLC vessel
5820 770.007
flat
16 215 × g
Ø 11 mm
14 000 rpm
1/12
7.4 cm
Cryo tube
5820 769.009
flat
18 188 × g
1.0 mL – 2.0 mL
Ø 13 mm
14 000 rpm
1/12
8.3 cm