4 comandi rtu modbus, Fc3 read holding register – VEGA VEGAFLEX 83 Modbus and Levelmaster protocol PFA insulated Manuale d'uso
Pagina 79

79
11 Appendice
VEGAFLEX 83 • Protocollo Modbus e LevelMaster
41838-IT
-130416
Register Name
Register Number
Type
Note
TV
2106
DWord
Third Variable in Byte Order ABCD DCBA (Little En-
dian)
QV
2108
DWord
Quarternary Variable in Byte Order DCBA (Little En-
dian)
Status
2200
DWord
See Register 100
PV
2202
DWord
Primary Variable in Byte Order BACD (Middle Endian)
SV
2204
DWord
Secondary Variable in Byte Order BACD (Middle En-
dian)
TV
2206
DWord
Third Variable in Byte Order BACD (Middle Endian)
QV
2208
DWord
Quarternary Variable in Byte Order BACD (Middle En-
dian)
Unit Codes for Register 104, 108, 112, 116
Unit Code
Measurement Unit
32
Degree Celsius
33
Degree Fahrenheit
40
US Gallon
41
Liters
42
Imperial Gallons
43
Cubic Meters
44
Feet
45
Meters
46
Barrels
47
Inches
48
Centimeters
49
Millimeters
111
Cubic Yards
112
Cubic Feet
113
Cubic Inches
11.4 Comandi RTU Modbus
FC3 Read Holding Register
Tramite questo comando è possibile leggere un numero qualsiasi (1-127) di holding register. Vengo-
no trasmessi il registro iniziale, a partire dal quale va effettuta la lettura, e il numero di registri.
Request:
Parameter
Length
Code/Data
Function Code
1 Byte
0x03
Start Address
2 Bytes
0x0000 to 0xFFFF