5 comandi levelmaster, Fc43 sub 14, read device identification – VEGA VEGABAR 82 Modbus and Levelmaster protocol - Operating Instructions Manuale d'uso
Pagina 76

76
10 Appendice
VEGABAR 82 • Protocollo Modbus e LevelMaster
46294-IT
-131205
Request:
Parametro
Length
Code/Data
Function Code
1 Byte
0x11
Response:
Parametro
Length
Code/Data
Function Code
1 Byte
0x11
Byte Number
1 Byte
Slave ID
1 Byte
Run Indicator Status
1 Byte
FC43 Sub 14, Read Device Identification
Con questo codice di funzione può essere richiesta la Device Identification.
Request:
Parametro
Length
Code/Data
Function Code
1 Byte
0x2B
MEI Type
1 Byte
0x0E
Read Device ID Code
1 Byte
0x01 to 0x04
Object ID
1 Byte
0x00 to 0xFF
Response:
Parametro
Length
Code/Data
Function Code
1 Byte
0x2B
MEI Type
1 Byte
0x0E
Read Device ID Code
1 Byte
0x01 to 0x04
Confirmity Level
1 Byte
0x01, 0x02, 0x03, 0x81, 0x82, 0x83
More follows
1 Byte
00/FF
Next Object ID
1 Byte
Object ID number
Number of Objects
1 Byte
List of Object ID
1 Byte
List of Object length
1 Byte
List of Object value
1 Byte
Depending on the Object ID
10.5 Comandi LevelMaster
Il VEGABAR 82 è idoneo anche ad essere collegato ai seguenti RTU con protocollo LevelMaster.
Questo protocollo viene spesso designato come "Protocollo Siemens" o "Protocollo serbatoio".
RTU
Protocol
ABB Totalflow
Levelmaster
Kimray DACC 2000/3000
Levelmaster